INSTRUCTIONS TO FIX LEAK IN VENOUS PORT OF CALIFIA PATIENT MODULE (CPM)

BACKGROUND:

Use these instructions to service a leak at the junction between the venous panel-mount quick connector and the venous valve threaded white bushing.

  1. PREPARATION:
    1. Drain the reservoir in its entirety
    1. Power off the unit
    1. Disconnect all external tubing
    1. Disconnect the power cable and Ethernet cable
  2. Move the Califia Patient Module (CPM) to a table or benchtop with plenty of workspaces
  3. REQUIRED TOOLS